Innovation in distillation testing
Automatic Distillation testing
by Orbis BV
STARDist - Automatic Distillation Unit
Excellent performance in D86 testing.
STARDist offers unique features unseen in any other instrument, such as: Optical Dry Point Testing,
PreScan for vol. & temp. normalizing, VOC Cold Trap for loss reduction and efficient condenser cooling.
Performance:
The ASTM D86 method is full of details that affect distillation results. For instance; sample start temperature, condenser
cleaning, vapor probe position, boiling stones, heater plate, distillation speed and ,many more.
Because the operator plays a critical role in getting these details right, STARDist offers Guided Setup; a step by step
support that not only explains the how but also why of each step. More seasoned STARDist users can skip Guided Setup,
depending on their user rights.
STARDist’s accurate temperature measurement, precise volume detection, early speed prediction and agile heater
control further guarantee excellent distillation behavior on important parameters such as IBP, distillation rate and FBP.
Even with difficult or unknown samples on the first test.
Flask Holder
Optical Dry Point
Volume Scanner
Guided Setup PreScan
Always correct Pt100 position No external dry point probe required
Accurate detection of sample volume
Sofware:
While the workflow feels easy and intuitive, the STARDist app offers complete distillation control and in-depth insight in
test results.
Modern ways of data handling such as “print pdf”, “create JSON/XML file” and “export to LIMS/Windows PC” make
STARDist easy to integrate in any laboratory data workflow.
The STARDist software is license-free, continuously supported by Orbis and available as a free download in the App Store.
Validation
Real time validation against 11 method-defined limits such
as Time till IBP, Condenser temperature, FBP Time-out.
Summary
Complete overview with quick
access to most relevant data and
actions such as print, export to
LIMS and Optimize.
Results
Create and report any custom point

test is long finished. Print results in
both extensive table and graph
Verification & Calibration
Easy-to-follow workflow for
quarterly verification and yearly
calibration procedures.
Methods ASTM D86, D1078, D850, E123, IP195, DIN51751, ISO 3405, GOST 2177, JIS K2254.
User interface iPad with STARDist app software. Available in App Store. App & updates are free of charge.
Operator
support
features
Guided Setup: easy to follow start up procedure to ensure all operators act the same (set mandatory on/off in user
settings)
PreScan: measures sample volume and temperature in the receiving chamber before each test. Actual measured
volume is then normalised to 100% volume, and receiving chamber temp. is automatically set to sample temp.
(provided whithin method limits).
Results validation: quick pass / fail test validation against method limits such as “time to IBP”, “FBP temp., etc.
Optimization: automatically optimizes program settings with fixed InHeat and Final Heat settings after first
distillation test.
Flask holder: ensures easy and correct flask positioning for both 125 ml and 200 ml flasks. The vapor probe’s height
position relative to the flask’s side arm is always correct thanks to the vapor probe design in combination with the
strict flask dimensions.
Heating control Smart multiparameter-based algorithm calculates and applies the required heater settings for IBP, FBP and (4-5
ml p/min.) speed control in real-time. Suitable for unknown samples and/or complicated blends such as E20. Speed
control range: 2-11 ml p/min.
Heating system Low mass/low voltage heater. 2 user adjustable fans for extra fast cooling after test.
Automatic heater lift with correct flask pressure & positioning. Automatic shut-off in case of fire.
Condenser
cooling
Solid state: based on Peltier elements with heat-pipe assisted heat sinks and silent fans for heat dissipation. No
liquids involved.
Condenser temperature can be increased during the distillation to deal with “light start / heavy end” products to
prevent both evaporation loss and product waxing in the condenser tube. Temperature range: 0 – 65°C. Resolution
0.1°C.
Volume
detection
Smart HD CCD camera for accurate volume measurement. Detects actual bottom of meniscus. Suitable for “smoke
producing” products. Automatic calibration of the camera against cylinder. Resolution: 0.01 ml, accuracy: 0.01 ml,
charge volume: 0-105%.
Receiving
chamber
Built-in drop deflector. Cooling: similar to condenser cooling. 1 sensor for receiver chamber temperature and 1
sensor for sample temperature. Automatic lift moves receiver cylinder up to be sealed at the top, where a small
hole ensures exposure to atmospheric pressure. Then; a tube laid along the condenser tube acts as “VOC Cold Trap,
preventing vapors from escaping the receiver (loss).Temperature range: 0 – 45°C. Resolution 0.1°C.
Residue & Loss Various options for residue measurement: automatic prediction, camera measurement, manual measurement and
preset value. Loss correction is automatically applied to temperature readings.
Vapor
temperature
measurement
PT-100 class A probe with 10-point calibration data storage and automatic probe ID detection. Calibration
certificate standard supplied. True dynamic simulation of ASTM 7 & 8 in-glass thermometer behavior (lag time and
emergent stem).
Range: 0 – 450°C ASTM / 0 – 500°C absolute. Units: ºC, ºF.
Dry point
detect.
Optical Dry Point: IR sensor for automatic dry point detection. No dry point probe required. (Conventional dp probe still
available)
Pressure Built-in pressure sensor. Automatic correction of temperature results to atmospheric pressure. Range: 70 to 110
kPa, res. 0.1 kPa.
Cetane index Built-in function for automatic calculation of cetane index.
Fire safety UV sensor for fire detection. Built-in fire extinguisher. N or CO2 supply required from lab (connection hose is
supplied).
System health Automatic quick system health check before every distillation run to ensure all components are in excellent state.
Connectivity 1 x RJ45 (LAN) for connection to LIMS, Windows PC, FTP server, Kiosk printer and/or connection between multiple
STARDist units.
Dimensions Dimensions: 40cm x 40cm x 63cm (WxDxH), weight: 40 kg
Accessories Kiosk printer
Voltage 100-240 VAC 50/60 Hz
Power 1200 W
Lab conditions Environment temperature 10°C – 30°C. Environment humidity up to 80 % at 30°C.
Technical specifications
No rights can be derived from this brochure. Orbis BV reserves the right to make changes any time.
Contact details:
De Regge 32 - 8253 PG Dronten
The Netherlands
Phone +31-321 382354
Fax +31-321 382357
www.orbisbv.com
OTHER ORBIS BV PRODUCTS:
AIRSTAR
- Cloud and Pour Point Testing
- Cold Filter Plugging Point Testing
Features:
ColdBlock: AirSTAR’s integrated cooling unit ( to -105ºC)
CFPP & CP/PP Heads: easy to use & exchangeable
Traditional Cloud, Pour and CFPP test methods
Innovation in distillation
STARDist Automatic Distillation Unit
Part No. 913021.
STARDist ordering information
ASTM D1078 / D850 pack Add-on pack for 913030
Part No. 919114. Includes:
25 mm ceran heater plate
32 mm ceran heater plate
Distillation flask 200 ml straight neck
Centering device for 200 ml flask
Accessories and consumables:
Part No. 919061 - Heater element for STARDist
Part No. 919123 - PT-100 vapor probe
Part No. 919120 - Centering device 125 ml flask
Part No. 919121 - Centering device for 200 ml flask
Part No. 919030 - Set of 5: Distillation flask 125 ml
Part No. 919031 - Set of 5: Distillation flask 200 ml
Part No. 919105 - 100 ml receiver with brass foot
Part No. 919106 - Set of 5: 100 ml receiver without foot
Part No. 11003 - Condenser stopper cap
Part No. 919033 - 25 mm ceran heater plate
Part No. 919034 - 32 mm ceran heater plate
Part No. 919035 - 38 mm ceran heater plate
Part No. 919036 - 50 mm ceran heater plate
Part No. 919081 - Boiling stones 25 gr.
Part No. 919039 - Condenser cleaning tool
Your local distributor:
Basic consumable kit For ASTM D86 testing
Part No. 913030. Includes:
PT-100 temp. probe including calibration report
Distillation flask 125 ml straight neck
Centering device for 125 ml flask
100 ml receiver with brass foot
38 mm ceran heater plate
50 mm ceran heater plate
Boiling stones 25 gr.
Condenser cleaning tool