Methods ASTM D86, D1078, D850, E123, IP195, DIN51751, ISO 3405, GOST 2177, JIS K2254.
User interface iPad with STARDist app software. Available in App Store. App & updates are free of charge.
Operator
support
features
Guided Setup: easy to follow start up procedure to ensure all operators act the same (set mandatory on/off in user
settings)
PreScan: measures sample volume and temperature in the receiving chamber before each test. Actual measured
volume is then normalised to 100% volume, and receiving chamber temp. is automatically set to sample temp.
(provided whithin method limits).
Results validation: quick pass / fail test validation against method limits such as “time to IBP”, “FBP temp.”, etc.
Optimization: automatically optimizes program settings with fixed InHeat and Final Heat settings after first
distillation test.
Flask holder: ensures easy and correct flask positioning for both 125 ml and 200 ml flasks. The vapor probe’s height
position relative to the flask’s side arm is always correct thanks to the vapor probe design in combination with the
strict flask dimensions.
Heating control Smart multiparameter-based algorithm calculates and applies the required heater settings for IBP, FBP and (4-5
ml p/min.) speed control in real-time. Suitable for unknown samples and/or complicated blends such as E20. Speed
control range: 2-11 ml p/min.
Heating system Low mass/low voltage heater. 2 user adjustable fans for extra fast cooling after test.
Automatic heater lift with correct flask pressure & positioning. Automatic shut-off in case of fire.
Condenser
cooling
Solid state: based on Peltier elements with heat-pipe assisted heat sinks and silent fans for heat dissipation. No
liquids involved.
Condenser temperature can be increased during the distillation to deal with “light start / heavy end” products to
prevent both evaporation loss and product waxing in the condenser tube. Temperature range: 0 – 65°C. Resolution
0.1°C.
Volume
detection
Smart HD CCD camera for accurate volume measurement. Detects actual bottom of meniscus. Suitable for “smoke
producing” products. Automatic calibration of the camera against cylinder. Resolution: 0.01 ml, accuracy: 0.01 ml,
charge volume: 0-105%.
Receiving
chamber
Built-in drop deflector. Cooling: similar to condenser cooling. 1 sensor for receiver chamber temperature and 1
sensor for sample temperature. Automatic lift moves receiver cylinder up to be sealed at the top, where a small
hole ensures exposure to atmospheric pressure. Then; a tube laid along the condenser tube acts as “VOC Cold Trap”,
preventing vapors from escaping the receiver (loss).Temperature range: 0 – 45°C. Resolution 0.1°C.
Residue & Loss Various options for residue measurement: automatic prediction, camera measurement, manual measurement and
preset value. Loss correction is automatically applied to temperature readings.
Vapor
temperature
measurement
PT-100 class A probe with 10-point calibration data storage and automatic probe ID detection. Calibration
certificate standard supplied. True dynamic simulation of ASTM 7 & 8 in-glass thermometer behavior (lag time and
emergent stem).
Range: 0 – 450°C ASTM / 0 – 500°C absolute. Units: ºC, ºF.
Dry point
detect.
Optical Dry Point: IR sensor for automatic dry point detection. No dry point probe required. (Conventional dp probe still
available)
Pressure Built-in pressure sensor. Automatic correction of temperature results to atmospheric pressure. Range: 70 to 110
kPa, res. 0.1 kPa.
Cetane index Built-in function for automatic calculation of cetane index.
Fire safety UV sensor for fire detection. Built-in fire extinguisher. N or CO2 supply required from lab (connection hose is
supplied).
System health Automatic quick system health check before every distillation run to ensure all components are in excellent state.
Connectivity 1 x RJ45 (LAN) for connection to LIMS, Windows PC, FTP server, Kiosk printer and/or connection between multiple
STARDist units.
Dimensions Dimensions: 40cm x 40cm x 63cm (WxDxH), weight: 40 kg
Accessories Kiosk printer
Voltage 100-240 VAC 50/60 Hz
Power 1200 W
Lab conditions Environment temperature 10°C – 30°C. Environment humidity up to 80 % at 30°C.
Technical specifications
No rights can be derived from this brochure. Orbis BV reserves the right to make changes any time.