Department of Defense
Module 4 Data Validation Guidelines: Data Validation Procedure for Organic Analysis by GC
March 2021
Page 31 of 40
7.2 Retention Time Windows
Retention Time Windows (RTWs) are critical to GC systems, and can have a direct impact on
compound identification. RTWs are established to compensate for minor shifts in absolute
retention times as a result of variability in sample injections, gas flows, or normal
chromatographic variability. The width of the RTW is carefully established to minimize the
occurrence of both false positive and false negative results and improve reproducibility.
RTWs are required to be established for both primary and secondary columns.
RTW width is required to be determined at method setup and after major maintenance.
RTWs are established by analyzing a minimum of three standards over a 72- hour period and
calculating the standard deviation (± 3 SD). As an alternative, the laboratory can establish a
0.03-minute retention time width, whichever is greater. RTW position is established once per
initial calibration and at the beginning of each analytical sequence. Position should be set
using the midpoint standard of the Initial Calibration curve when an Initial Calibration is
performed. On days when the Initial Calibration is not performed, the retention times
displayed in the initial CCV are used to set the position.
Multicomponent analytes such as Technical Chlordane, Toxaphene, Strobane, Aroclors and
Fuels are distinguished based on the ranges of retention times for characteristic components.
Separate calibration standards are necessary for each multi-component target analyte like
Toxaphene. Typically, an alkane standard containing a homologous series of n-alkanes is
used for establishing retention times for fuels like diesel. When the whole response (including
the hump-o-gram) is used for quantitation, appropriate RTWs are established for the boiling
point range or carbon number range used to define each multicomponent analyte. According
to 8015, two specific gasoline components are used to establish the GRO range (2-
methylpentane and 1,2,4-trimethylbenzene) or the DRO range (C10 and C 28 alkanes).
Refer to the method or the QAPP for specific requirements. Retention time windows are
established similarly as described above for the two components that bracket the multi-
component analyte. The retention time range is then calculated based on the lower limit of
the RTW for the first eluting component and the upper limit of the RTW for the last eluting
component. When three to five unique peaks are used in the identification of Technical
Chlordane, Toxaphene, Strobane, and/or Aroclors, these individual quantitation peaks shall
have established RTWs as described above.
Evaluation of Retention Time Windows
Verify from the raw data that the RTWs were established and are calculated correctly. The
CCV can be used to verify RTWs, and at least two analytes per calibration curve should be
recalculated (if a standard deviation was used) to verify a correct window width and
placement. If deviations are discovered, the laboratory should be contacted to correct and re-
produce the report.
Verify that all reported analytes and surrogates are within their established RTWs for both the
primary and secondary column. If positive results (detects) were reported from peaks outside
of the established RTW, contact the laboratory to evaluate for false positives and re-issue a
revised report.
Retention times of the peaks for detects must fall within the calculated window for both
chromatographic columns, and the pattern of the peaks should match the pattern in the
CCVs on both columns for multicomponent analytes. If the peaks fall outside of the RTW on